• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK AXS CATALYST 7 .068IN ID X 132CM - US; CATHETER, THROMBUS RETRIEVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK AXS CATALYST 7 .068IN ID X 132CM - US; CATHETER, THROMBUS RETRIEVER Back to Search Results
Model Number IC068132
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/09/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the procedure, after aspiration, when the physician removed the subject catheter from the patient's anatomy, the proximal part of the subject catheter was found broken.The physician removed the fractured fragment of the subject catheter out of the patient's anatomy.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
H3 device evaluated by mfg: updated.H3 summary attached: updated.H4 manufacturing date: added.D4 expiration date: added.Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.During visual inspection, the catheter shaft was seen to be broken 6cm and 31cm from the catheter hub.The catheter shaft was seen to be kinked at 58cm and 60cm from the hub.There was dried blood seen along the distal shaft.Functional inspection was not required as the defect was confirmed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The catheter was seen to be broken therefore confirming the as reported event.The catheter was also seen to be kinked bent.The as reported 'catheter shaft broken/fractured during use', as well as the as analyzed 'catheter shaft broken/fractured during use', 'catheter shaft kinked/bent', will be assigned procedural factors as this complaint appears to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural factors during use.
 
Event or Problem Description
It was reported that during the procedure, after aspiration, when the physician removed the subject catheter from the patient's anatomy, the proximal part of the subject catheter was found broken.The physician removed the fractured fragment of the subject catheter out of the patient's anatomy.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXS CATALYST 7 .068IN ID X 132CM - US
Common Device Name
CATHETER, THROMBUS RETRIEVER
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key13684510
Report Number3008881809-2022-00108
Device Sequence Number11239296
Product Code NRY
UDI-Device Identifier07613327417227
UDI-Public07613327417227
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K183464
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/12/2023
Device Model NumberIC068132
Device Catalogue NumberIC068132
Device Lot Number23195781
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/09/2022
Supplement Date Received by Manufacturer05/16/2022
Initial Report FDA Received Date03/07/2022
Supplement Report FDA Received Date06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FLOWGATE BALLOON GUIDE CATHETER (STRYKER); SYNCHRO 14 GUIDEWIRE (STRYKER)
Patient SexUnknown
-
-