| Model Number |
TFAT30 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
12/16/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A physician reported that following an intraocular lens (iol) implant procedure, the patient experienced blurry vision.The iol was exchanged in a secondary procedure following the initial procedure.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
A sample device was not returned for analysis.Product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There has been two other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
Additional information has been requested and provided it was reported that a tfat30.180 lens was exchanged for a tfat40.180 lens due to blurry vision, mechanical complication of lens.This does not meet criteria for reporting based on applicable medical device regulations.A lens exchange from one power to a different power is an attempt to adjust or fine tune the refractive outcome trying to achieve a desired target when the initial lens calculations did not achieve the intended outcome.There is no reported defect or fault with the lens.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on information received following submission of the initial report, this event does not meet criteria for reporting as a serious injury.The manufacturer internal reference number is: (b)(4).
|
| |
|