Model Number C111
Medical Device Problem Code Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The last calibration performed on (b)(6) 2021 was ok.
 
Event or Problem Description
The initial reporter stated they received discrepant results for one patient sample tested with ggt-2-glutamyl transferase ver.2 standardized against ifcc / szasz on a cobas c 111.The sample initially resulted in a ggt value of 362.4 u/l and this value was reported outside of the laboratory.The sample was repeated on (b)(6) 2022, resulting in a value of 9.9 u/l.The repeat value agreed with the patient's clinical picture.The ggt reagent lot number was 530461-01.The reagent expiration date was requested, but not provided.
 
Additional Manufacturer Narrative
Quality control data for one level was provided for dates 07-feb-2022 to 13-feb-2022; the results were ok.Upon review of analyzer data, there was no indication of a technical issue.The investigation could not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS C111
Common Device Name
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE INSTRUMENT CENTER AG TEGIMENTA
forrenstrasse
na
rotkreuz 6343
SZ   6343
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key13685033
Report Number1823260-2022-00591
Device Sequence Number5110768
Product Code JJE
UDI-Device Identifier04015630924769
UDI-Public04015630924769
Combination Product (Y/N)N
Initial Reporter CountryHR
PMA/510(K) Number
K071211
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberC111
Device Catalogue Number04777433001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/10/2022
Supplement Date Received by Manufacturer03/22/2022
Initial Report FDA Received Date03/07/2022
Supplement Report FDA Received Date04/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-