| Model Number |
C111 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The last calibration performed on (b)(6) 2021 was ok.
|
| |
|
Event or Problem Description
|
|
The initial reporter stated they received discrepant results for one patient sample tested with ggt-2-glutamyl transferase ver.2 standardized against ifcc / szasz on a cobas c 111.The sample initially resulted in a ggt value of 362.4 u/l and this value was reported outside of the laboratory.The sample was repeated on (b)(6) 2022, resulting in a value of 9.9 u/l.The repeat value agreed with the patient's clinical picture.The ggt reagent lot number was 530461-01.The reagent expiration date was requested, but not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Quality control data for one level was provided for dates 07-feb-2022 to 13-feb-2022; the results were ok.Upon review of analyzer data, there was no indication of a technical issue.The investigation could not identify a product problem.The cause of the event could not be determined.
|
| |
|