| Model Number |
1054655 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/08/2014
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This mdr is being submitted as part of a batch submission of previously closed complaints that were reassessed as reportable foam degradation complaints; discovered as part of a retrospective remediation review.
|
| |
|
Event or Problem Description
|
|
A ventilator was returned to the manufacturer for routine preventative maintenance.There was no allegation of device malfunction.The device was not in patient use.During the evaluation of the device at the manufacturer's service center, the air path foam, stirring fan, and flow sensor assembly were replaced due to preventative maintenance.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer previously reported the investigation was ongoing and a follow up report would be submitted when the investigation was complete.No follow up report is required.
|
| |
|
Search Alerts/Recalls
|