Concomitant medical products: other relevant device(s) are: product id: evproplus-29us, serial/lot #: (b)(4), ubd: 01-oct-2023, udi#: (b)(4); product analysis: the valve remains implanted and the dcs was discarded, therefore no product analysis can be performed.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
|
Medtronic received information that during full release of this transcatheter bioprosthetic valve, the valve dislodged.It was reported that the possible cause of the dislodgement was the weak forward tension applied to the delivery catheter system (dcs), and that the implant depth became shallower at the time of full release.A second transcatheter bioprosthetic valve was implanted.No additional adverse patient effects were reported.
|