| Model Number |
C111 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Na.
|
| |
|
Event or Problem Description
|
|
The initial reporter stated they received discrepant results for one patient sample tested with crej2 creatinine jaffé gen.2 (compensated) on a cobas c 111.No incorrect results were reported outside of the laboratory.The sample initially resulted in a creatinine value of 2.68 mg/dl and repeated as 0.44 mg/dl.The creatinine reagent lot number was 57929101, with an expiration date of 30-jun-2023.
|
| |
|
Additional Manufacturer Narrative
|
|
A general reagent issue could be ruled out because calibration and quality controls were acceptable.The customer cleans the sample probe every 7 to 10 days.Per product labeling, the probe has to be cleaned manually on a weekly basis and deproteinized daily.The issue is consistent with a failed pipetting on the analyzer related to inadequate customer maintenance.
|
| |
|