| Model Number |
K173182 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Unspecified Musculoskeletal problem (4535)
|
| Date of Event |
02/08/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Based on the reported event, it seems as though the implant was not expanded during the initial centerline/prolift surgery.This would have essentially left the prolift acting like an undersized static interbody and would have allowed for migration and wobble between vertebrae.
|
| |
|
Event or Problem Description
|
|
Centerline amd prolift removal, it was reported that the "screw placement and cage wee backing out, cage not expanded fully, no issue removing, moved up a level.".
|
| |
|