• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIFE SPINE, INC. PROLIFT; INTERVERTEBRAL BODY FUSION DEVICE/

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIFE SPINE, INC. PROLIFT; INTERVERTEBRAL BODY FUSION DEVICE/ Back to Search Results
Model Number K173182
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Unspecified Musculoskeletal problem (4535)
Date of Event 02/08/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the reported event, it seems as though the implant was not expanded during the initial centerline/prolift surgery.This would have essentially left the prolift acting like an undersized static interbody and would have allowed for migration and wobble between vertebrae.
 
Event or Problem Description
Centerline amd prolift removal, it was reported that the "screw placement and cage wee backing out, cage not expanded fully, no issue removing, moved up a level.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROLIFT
Common Device Name
INTERVERTEBRAL BODY FUSION DEVICE/
Manufacturer (Section D)
LIFE SPINE, INC.
13951 s. quality drive
huntley IL 60142
Manufacturer (Section G)
LIFE SPINE, INC.
13951 s. quality drive
huntley IL 60142
Manufacturer Contact
angela batker
13951 south quality drive
huntley, IL 60142
8478846117
MDR Report Key13685466
Report Number3004499989-2022-00005
Device Sequence Number13938973
Product Code MAX
UDI-Device Identifier00190837022730
UDI-Public00190837022730
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K173182
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberK173182
Device Catalogue Number58-1228-0813D
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/10/2022
Initial Report FDA Received Date03/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/21/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-