| Model Number |
1012445-15 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Material Rupture (1546); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported complaints appear to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a 75% stenosed lesion in the distal right coronary artery (drca) with moderate calcification and mild tortuosity.A 2.00 x 15mm nc trek balloon dilatation catheter (bdc) was advance to the target lesion; however resistance was met with the anatomy.During inflation the balloon was inflated three times for 2-3 seconds each time; however the balloon ruptured on the third inflation at 8 atmospheres.The bdc was removed from the patient with resistance due to anatomical challenges.A non-abbott balloon was used to complete the procedure.There were no adverse patient effects reported or clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|