|
The following sections were updated in follow-up 1.One 13.5f guidestar steerable guiding sheath was received without the dilator.There were no other accessories.Traces of blood were found on and inside the sheath.According to the event description summary, the sheath was leaking due to crack on the distal part of the handle.A visual inspection of the handle showed no signs of breakage or damage.When the deflection collar was rotated, the sheath deflected as intended.The valve was viewed under a 10x microscope and looked normal with no anomalies.The sheath was manually leak tested with the syringe connected to the stopcock and the tip occluded.The sheath did not exhibit leakage when tested using this method.The sheath was then tested using the iso 11070 for liquid leakage test method.The device was occluded at the distal tip and pressurized to 38kpa and held at this pressure for 30 seconds and no leakage was observed from the hemostatic valve.The sheath was further pressurized beyond 300kpa and no leakage was observed from the stopcock, sideport tubing or through the handle.The device history records were reviewed to confirm that the device passed all applicable in-process and final inspections.Returned device analysis revealed the sheath was within manufacturing specifications.According to the event description summary, the sheath was leaking due to crack on the distal part of the handle.Returned device analysis revealed the handle of the sheath was intact and showed no signs of breakage or damage.When the deflection collar was rotated, the sheath deflected as intended.The sheath did not leak when tested manually.The sheath also met the iso leakage test method requirement.According to the dhr, the sheath passed all in-process and final inspection steps including visual, mechanical, dimensional and leak testing.No manufacturing defects were found.Per qa procedure (destino steerable guiding sheath in-process and final inspection) handle assembly: verify the two sides of the handle have minimal gaps.If the gap affects the function of the handle, reject the unit.Perform deflection test according to procedure.The deflection test is performed by manufacturing personnel at 100% and inspected by quality assurance personnel.Aql sample size: ansi z 1.4, gen level i, normal, aql 1.5 normal.Leak test: sample size: ansi z 1.4, gen level i, normal, aql 1.5 normal.Perform leak test according to procedures based on available leak tester.The leak test is performed by manufacturing personnel at 100% and is observed by quality assurance personnel at aql level ansi z 1.4, gen level i, normal, aql 1.5 normal.Per ifu the steerable sheath must be thoroughly flushed with either saline or heparinized saline and free of air prior to use to avoid air embolism to the patient.Aspiration and flushing of the sheath should be performed frequently to help minimize the potential for air embolism.For injecting or aspirating through the sheath, use the sideport only with stopcock.Prior to infusion, remove all air using the sideport.No further follow-up is required.Based on the investigation, a capa is not required as findings did not identify a design, labeling or manufacturing non-conformity.In addition, there was no new failure mode identified and the risk remains acceptable.The event will be re-evaluated if additional information becomes available.Oscor will continue to monitor this event type and risk.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
|