| Model Number |
151P |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/31/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received information alleging an issue related to a remstar device's sound abatement foam.There was no report of harm or injury.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
|
| |
|
Additional Manufacturer Narrative
|
|
This mdr is being submitted as part of a batch submission of complaints that were reassessed as reportable foam degradation complaints; discovered as part of a retrospective remediation review.
|
| |
|
Search Alerts/Recalls
|