Model Number BA25-120/I20-40
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Codes Aneurysm (1708); Failure of Implant (1924)
Date of Event 03/02/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx with duraply.
 
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with implant of an afx bifurcated stent graft and an afx vela suprarenal.Approximately six (6) years post initial procedure, aneurysm enlargement with a type ib endoleak and an endoleak at an indeterminate origin were identified.Reintervention was completed with unknown non-endologix devices.The patient status post-reintervention was not reported.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the type ib and type iv endoleaks, sac growth, and additional endovascular procedure are unconfirmed.This is not consistent with the reported adverse event/incident.Procedure related harms, device, user, procedure or anatomy relatedness of this complaint could not be determined with the medical records available for review.The final patient status remains unknown.The final patient status was not made available to endologix.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx with duraply.H6: investigation finding codes - remove code 3233.H6: investigation conclusion codes - remove code 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key13687732
Report Number2031527-2022-00042
Device Sequence Number5110844
Product Code MIH
UDI-Device Identifier00818009011985
UDI-Public(01)00818009011985(17)170923
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 03/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/23/2017
Device Model NumberBA25-120/I20-40
Device Lot Number1278499014
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/02/2022
Supplement Date Received by Manufacturer04/25/2022
Initial Report FDA Received Date03/08/2022
Supplement Report FDA Received Date05/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/14/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX VELA SUPRARENAL (LN 1286549044)
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR
Patient SexMale
-
-