DEPUY SYNTHES PRODUCTS LLC 1.85MM X 16 MM FRTR,CRNA; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Model Number A-CRN-G1
Medical Device Problem Codes Dull, Blunt (2407); Separation Problem (4043)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/17/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.  quality engineering evaluated the device and it was determined that the reported condition of the cutter device being worn, identified during service and evaluation was confirmed.The assignable root cause of the worn cutter was determined to be traced to component failure from wear and the dirty condition of the device was determined to be traced to environment.Udi: (b)(4).
 
Event or Problem Description
It was reported that during in-house engineering evaluation, it was determined that the router of the cutter device was generally worn, dirty and not cleaned properly.It was noted in the service order that the handpiece device was overheating and the cutter device was unable to be detached.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.85MM X 16 MM FRTR,CRNA
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kara ditty-bovard
4500 riverside drive
palm beach gardens, FL 33410
6107428552
MDR Report Key13688145
Report Number1045834-2022-00285
Device Sequence Number4930237
Product Code HBE
UDI-Device Identifier00845384012603
UDI-Public00845384012603
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K113476
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/26/2026
Device Model NumberA-CRN-G1
Device Catalogue NumberA-CRN-G1
Device Lot NumberR233146579
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2022
Initial Report FDA Received Date03/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-