MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PINPOINT LAPAROSCOPE - 5MM, STANDARD, 30 WITH WOLF AND STORZ ADAPTOR; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PINPOINT LAPAROSCOPE - 5MM, STANDARD, 30 WITH WOLF AND STORZ ADAPTOR; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number SC9534
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that there was blurry image.Please note that the procedure was completed successfully.
 
Event or Problem Description
It was reported that there was blurry image.Please note that the procedure was completed successfully.
 
Additional Manufacturer Narrative
Reported failure mode: "blurry & not clear image /half black image".Probable root cause: the reported failure mode was confirmed and probable root cause was likely due to damage during use/handling of the endoscope.The device was returned for investigation and the reported failure mode was confirmed.The reported failure mode will be monitored for future re occurrence.For information and warnings related to the proper use and maintenance of this device, please reference and adhere to information presented in the associated product ifu.Manufacturing date is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINPOINT LAPAROSCOPE - 5MM, STANDARD, 30 WITH WOLF AND STORZ ADAPTOR
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
burnaby
san jose CA 95138
Manufacturer Contact
chester rebugio
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key13688717
Report Number3012345110-2022-00001
Device Sequence Number5308988
Product Code GCJ
UDI-Device Identifier00858701006131
UDI-Public00858701006131
Combination Product (Y/N)N
Initial Reporter CountryKU
PMA/510(K) Number
K161792
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSC9534
Device Catalogue NumberSC9534
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/07/2022
Supplement Date Received by Manufacturer02/07/2022
Initial Report FDA Received Date03/08/2022
Supplement Report FDA Received Date04/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-