MAUDE Adverse Event Report: LAKE REGION MEDICAL PREFORM GUIDEWIRE-SAFARI2; WIRE GUIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LAKE REGION MEDICAL PREFORM GUIDEWIRE-SAFARI2; WIRE GUIDE Back to Search Results
Model Number H74939406S1
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This medwatch report is being filed based on images received on february 22, 2022, which present a fracture of the core wire.The device was not received for analysis at the time of this report; therefore, no physical analysis of the product could be performed.A review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.During manufacturing and packaging of this product the operators are instructed to 100% visually inspect for any obvious defect prior to shipment.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.The supplied images presented a fracture of the core wire and stretched coil wraps in the distal tip extending proximally from the j-straightener/introducer.As described in the device instructions for use (dfu): ensure a catheter is properly placed in the ventricle or other intended treatment area.Carefully remove the safari2tm guidewire from the hoop by grasping the proximal end of the j-straightener separating the straightener from the hoop.After inspection of the safari2tm guidewire, advance the j-straightener over the distal section of the safari2tm guidewire to straighten the curve.Insert the safari2tm guidewire into the hub of the catheter with the aid of the j-straightener.Carefully advance the distal tip of the safari2tm guidewire into the lumen of the catheter.Remove the safari2tm guidewire j-straightener by withdrawing it over the length of the safari2tm guidewire.Advance the safari2tm guidewire through the catheter under fluoroscopic guidance.As indicated in the dfu warnings: "do not torque this guidewire." at this time, it is not possible to assign a definitive root cause for the event as reported.Based on the information provided to date, it appears that clinical and/or procedural factors may have impacted on the event as reported.If any further relevant information is provided, a follow up medwatch report will be submitted.
 
Event or Problem Description
The wire was inserted into the introducer in the groin, so was inside the patient at the time of the event, but may or may not have had vessel wall contact.No vascular damage noted and the wire was removed and a new wire from the same lot was used to complete the procedure.No patient complications.Was issue present upon opening package?: no.Photo or video of issue: yes.Question: are any patient status updates available? answer: no.Question: describe anatomy (bicuspid, gothic arch, acute bend, horizonal aorta, degree of curvature etc)? answer: normal anatomy.Question: is it known what caused the device damage? answer: unknown.Question: where specifically on the device did the damage occur? answer: when the j-straightener was pulled back the wire uncoiled.Additional information provided via email on february 23, 2022: you note that the wire was inserted in the introducer in the groin.With this in mind, when exactly was this damage noticed upon insertion? the damage to the wire was noted when the j-straightener was pulled back , this after a few cm of insertion of the wire.A.At what point in the procedure was the safari2 removed from the patient? the damaged safari wire was removed during the initial period of the procedure where access to the groin vessels have been obtained with 6-8 fr.Introducers.
 
Additional Manufacturer Narrative
Device evaluation: as received, the specimen consisted of one-1 each safari2 275cm sml crv; returned coiled, loose with the distal tip lodged within the j-straightener and double-bagged within "zip-lock" style poly biohazard pouches.A review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.During manufacturing and packaging of this product the operators are instructed to 100% visually inspect for any obvious defect prior to shipment.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.The specimen presented a ductile, tensile overload of the core wire an unknown distance from the distal tip weld mass with concurrent stretched coil damage; the distal tip is lodged within the lumen the j-straightener attachment 3.40cm from the distal tip of the j-straightener.Approximately 1.80cm of the metallic core wire proximal of the fracture is protruding through the coil wraps.The exposed core wire presented bend damage 0.60 to 0.754cm from the proximal aspect of the wire fracture.Our investigation was unable to confirm that the product did not meet specification prior to shipment.The investigation concluded that the product met specification at the time of shipment.As described in the dfu instructions for use (dfu): 1.Ensure a catheter is properly placed in the ventricle or other intended treatment area.2.Carefully remove the safari2tm guidewire from the hoop by grasping the proximal end of the j-straightener separating the straightener from the hoop.3.After inspection of the safari2tm guidewire, advance the j-straightener over the distal section of the safari2tm guidewire to straighten the curve.4.Insert the safari2tm guidewire into the hub of the catheter with the aid of the j-straightener.5.Carefully advance the distal tip of the safari2tm guidewire into the lumen of the catheter.6.Remove the safari2tm guidewire j-straightener by withdrawing it over the length of the safari2tm guidewire.7.Advance the safari2tm guidewire through the catheter under fluoroscopic guidance.At this time, it is not possible to assign a definitive root cause for the event as reported.Based on the information provided to date, it appears that handling and/or procedural factors have contributed to the event as reported.If any further relevant information is received, a follow up medwatch report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PREFORM GUIDEWIRE-SAFARI2
Common Device Name
WIRE GUIDE
Manufacturer (Section D)
LAKE REGION MEDICAL
340 lake hazeltine drive
chaska MN 55318
Manufacturer (Section G)
LAKE REGION MEDICAL
340 lake hazeltine drive
chaska MN 55318
Manufacturer Contact
sharon seifert
340 lake hazeltine drive
chaska, MN 55318
9526418518
MDR Report Key13688838
Report Number2126666-2022-00013
Device Sequence Number1945010
Product Code DQX
UDI-Device Identifier08714729887591
UDI-Public08714729887591
Combination Product (Y/N)N
Initial Reporter CountryDA
PMA/510(K) Number
K151244
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/24/2024
Device Model NumberH74939406S1
Device Catalogue Number39406S
Device Lot Number6618383
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/22/2022
Supplement Date Received by Manufacturer03/22/2022
Initial Report FDA Received Date03/08/2022
Supplement Report FDA Received Date03/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age84 YR
Patient SexMale
-
-