Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 02/15/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient had a trabecular metal patella.It peg sheared off from the patella approximately one year four months post procedure.The patient was revised with a cemented patella.No additional information according to per.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.  performed a visual inspection of the returned patella and found it to exhibit signs of being implanted and the post has fractured/sheared off.Sem analysis: fracture surface on one of the regions of the peg sample revealed artifacts that are suspected to be fatigue striations, which are a tell-tale sign for fatigue mode of failure.The patella peg sample, shows the fractured region and suspected fatigue striations.The device history records were reviewed for deviations and/or anomalies with no related anomalies/deviations identified.Insufficient information provided to perform a compatibility check.Review of complaint history found no additional related issues for the reported part and lot combination.Medical records/radiographs were not provided.Root cause of the fatigue fracture cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Additional Manufacturer Narrative
 this follow-up report is being submitted to relay additional information.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEXGEN COMPLETE KNEE TRABECULAR METAL STANDARD PRIMARY PATELLA
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER TMT
10 pomeroy road
parsippany NJ 07054
Manufacturer (Section G)
ZIMMER TMT
10 pomeroy road
parsippany NJ 07054
Manufacturer Contact
jennifer rapsavage
10 pomeroy road
parsippany, NJ 07054
5745260384
MDR Report Key13689085
Report Number3005751028-2022-00006
Device Sequence Number1905744
Product Code JWH
UDI-Device Identifier00889024309838
UDI-Public(01)00889024309838(17)240930(10)64386857
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Model NumberN/A
Device Catalogue Number00587806541
Device Lot Number64386857
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/15/2022
Supplement Date Received by Manufacturer05/09/2022
06/29/2022
Initial Report FDA Received Date03/08/2022
Supplement Report FDA Received Date06/02/2022
06/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/17/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
Patient Weight125 KG
-
-