| Model Number |
7211006 |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Excess Flow or Over-Infusion (1311)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: case (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during surgery, the dyonics 25 pump and/or dyonics 25 tubing ran fluid continuously and did not sense back pressure.It filled the patients quad full of fluid.The surgeon was able to catch the issue and stop the pump.It is unknown if a back-up device was available and if a delay occurred.No patient complications occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).510k no.Updated to unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that during surgery, the 'dyonics 25 pump and/or dyonics 25 tubing' ran fluid continuously and did not sense back pressure.It filled the patient's quadriceps full of fluid.The surgeon was able to catch the issue and stop the pump.It is unknown if a back-up device was available and if a surgical delay took place.No further complications occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
H10, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.Insufficient product identification information was provided and thus a manufacturing record review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.Insufficient product identification information was provided and thus a complaint history review could not be conducted.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.Insufficient product identification information was provided and thus a risk management review could not be conducted.A clinical review states that based on insufficient information, a thorough medical assessment could not be performed at this.Should any additional relevant clinical information be provided, this complaint would be re-assessed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
| |
|