| Model Number |
2076-6020 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The angiosulpt device was used to treat a peripheral lesion.During the first inflation below rbp, the balloon did not inflate.Upon removal outside of the patient's body, a leak from the tip was observed, but could not confirm exact location of the leak.The procedure was completed with a new angiosculpt.No patient injury reported.This product problem is being submitted conservatively because the balloon may have ruptured below rbp.
|
| |
|
Additional Manufacturer Narrative
|
|
The patient's dob or age at time of event, gender, weight, ethnicity, and race are unknown.This information was not available from the facility.Patient information regarding relevant tests/laboratory data or medical history are unknown.This information was not available from the facility.Attempts to obtain the information has not been successful.Foreign: (b)(6).The angiosculpt device has not been returned by the facility, thus no returned product investigation was performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Block b1/b5/h1: based on the returned product evaluation, this is no longer reportable for a product problem.Block d9/g3: the angiosculpt device was returned for evaluation.Block h3: visual inspection found a bent distal tip, a damaged balloon with the distal end inverted, and a deformed scoring element.During functional testing, using an indeflator filled with green color dye water, the device was pressurized to nominal (8 atm) and a hub leak was present at the guidewire port.The balloon was able to inflate with a misaligned scoring element noted.No balloon rupture was observed.Block h6: a balloon rupture could not be confirmed.However, a hub leak was noted, but unable to determine the root cause of the observed failure.A review of the dhr and manufacturing process could not confirm a cause related to design and/or process failure.
|
| |
|
Event or Problem Description
|
|
This product problem is no longer reportable because no balloon rupture was observed during the returned device evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: component code corrected from 419 (balloon) to 4761 (access port) based on the returned device evaluation stated in supplemental report #1.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|
| |
|