SMITH & NEPHEW, INC. DYONICS 25 INFLW/OUTFLW FORK SUCTN (3); ARTHROSCOPE Back to Search Results
Model Number 7211006
Medical Device Problem Code Increase in Pressure (1491)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: case (b)(4).
 
Event or Problem Description
It was reported that during use in a shoulder arthroscopy, the "dyonics 25 inflow/outflow fork suction" was over pressurizing the shoulder joint.The procedure was successfully completed without surgical delay using the same device.No patient complications were reported.
 
Additional Manufacturer Narrative
H10: h3, h6: the reported device was received for evaluation.A visual inspection of the exterior of the device did not reveal any issues but after disassembly of the cassette, rusty bearing was found in the impeller.A functional evaluation revealed that the impeller was noisy.There was no issue with the out pressure.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause of the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.Corrected data: a1 should be read in blank.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYONICS 25 INFLW/OUTFLW FORK SUCTN (3)
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13690266
Report Number1643264-2022-00077
Device Sequence Number1943420
Product Code HRX
UDI-Device Identifier03596010543592
UDI-Public03596010543592
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K051326
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Model Number7211006
Device Catalogue Number7211006
Device Lot Number50976089
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/15/2022
Supplement Date Received by Manufacturer10/19/2023
Initial Report FDA Received Date03/08/2022
Supplement Report FDA Received Date10/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-