• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. PRECICE SYSTEM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. PRECICE SYSTEM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number P10.7-80B275
Medical Device Problem Codes Mechanical Problem (1384); Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/29/2022
Type of Reportable Event Malfunction
Event or Problem Description
Information was received that the nail stopped during the lengthening protocol.Per the reporter, the nail was moving inside the body part.No patient adverse event was reported.
 
Additional Manufacturer Narrative
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
 
Event or Problem Description
Additional information was received that "the thinner part is going inside the thicker part.It is loose".
 
Additional Manufacturer Narrative
H6 (device code 4003 was previously submitted.The reporter has confirmed this event was not a migration issue, but a mechanical issue).
 
Additional Manufacturer Narrative
Additional data: b5, h3, h6, h10.Device evaluation: upon return, visual inspection of the nail revealed no external damage.X-ray images of the internal components showed no damage and revealed the nail to have been partially distracted.The nail was functionally tested and was able to distract and retract without any abnormal noise using the high-speed magnet tool and the external remote controller (erc).The length as received was measured at 274.49 mm.The maximum and minimum length were measured at 355.57 mm and 273.12 mm respectively; giving a total stroke measurement of 82.45 mm, which meets the specification.Distraction force testing was measured at 292.40 lbs.Which meets the specification.The reported failure could not be confirmed as the nail was found to be functional and met the specifications.As part of this complaint investigation, there were no loose components found within the nail.Device records review: review of the device history records indicated, the unit met all required quality specifications and testing prior to product departure.
 
Event or Problem Description
No additional information has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECICE SYSTEM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr ste 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr ste 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr ste 100
aliso viejo, CA 92656
MDR Report Key13690369
Report Number3006179046-2022-00136
Device Sequence Number17395635
Product Code HSB
UDI-Device Identifier00812258020262
UDI-Public812258020262
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K191336
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberP10.7-80B275
Device Lot Number1061102AAA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2022
Supplement Date Received by Manufacturer03/09/2022
04/14/2022
Initial Report FDA Received Date03/08/2022
Supplement Report FDA Received Date03/11/2022
04/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age28 YR
Patient SexFemale
Patient Weight47 KG
-
-