| Model Number |
133S-MX-14-T |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
02/08/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Company representative reported "punctured" against unknown side device.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: crease flat ,yellow particle in the shell and device appearance (no openings was found in the device).Leak test was performed, and no leak was found.A microscopic analysis was performed which no identify opening.No openings was found in the device.Based on the device analysis the final assessment is: no issues found related with the manufacturing process.Additional, changed, and/or corrected data: d9, h3, h6.
|
| |
|
Event or Problem Description
|
|
Company representative reported "punctured" against unknown side device.The device has been explanted.
|
| |
|