BOSTON SCIENTIFIC CORPORATION ORISE PROKNIFE; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number M00519361
Medical Device Problem Codes Mechanical Problem (1384); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medical device problem code: (b)(4).Investigation results the returned orise proknife was analyzed, and a visual analysis found no problems with the handle, however a kink was observed in the working length.Additionally, the tip of the electrode was observed to be detached.The ceramic tip was removed from the device to observe the location of detached electrode.A functional evaluation noted that no resistance was felt when actuating the thumb slide on the handle, however the electrode did not extend from the tip.No other issues were noted.Based on all available information and the condition of the returned device, it is likely that procedural factors, such as the length of time used, power settings required, handling technique, and/or tissue composition resulted in the electrode detachment.The investigation concluded the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications at the time of release for distribution.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Event or Problem Description
It was reported to boston scientific corporation that a 2 mm orise proknife electrode was used during an endoscopic submucosal dissection (esd) procedure performed on (b)(6) 2022.During the procedure, the electrode does not move at all.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.Note: this event has been deemed an mdr-reportable event based on investigation results which revealed that the electrode was detached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORISE PROKNIFE
Common Device Name
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road cork
cork IRELA ND
EI   IRELAND
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13690727
Report Number3005099803-2022-01006
Device Sequence Number1906315
Product Code KNS
UDI-Device Identifier08714729999898
UDI-Public08714729999898
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K200404
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00519361
Device Catalogue Number1936
Device Lot Number27799549
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/04/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/10/2022
Initial Report FDA Received Date03/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-