• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, MULTIFOCAL INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, MULTIFOCAL INTRAOCULAR Back to Search Results
Model Number DFW150
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Halo (2227); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 01/17/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for analysis.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that after the intraocular lens (iol) was implanted in the patient's right eye, the patient vision was still very blurry with lots of halos.The patient had the iol explanted.Another johnson and johnson iol (same model, different diopter, 20.5) was implanted as a replacement.There was no medical or surgical interventions and no patient injury.Planned interventions include follow-up with meibomian gland dysfunction od, epiretinal membrane (erm) od, band keratopathy od, and pseudophakia od.The patient's vision is clearing up and there is no pain.No further information was provided.
 
Additional Manufacturer Narrative
Corrected data: in review, it was noted that in the initial mdr report, it was inadvertently indicated that the replacement lens was of "20.5" diopter size which is incorrect.The information has been corrected in this supplemental mdr accordingly: section b5: another johnson and johnson iol (same model, different diopter, 21.0) was implanted as a replacement.Additional information: section d9: device available for evaluation? yes.Section d9: returned to manufacturer: yes.Section d9: date returned to manufacturer: mar 10, 2022.Section h3: evaluated by manufacturer: yes.Device evaluation: visual inspection under magnification revealed that the lens was received cut in half.The lens was cleaned, and no issues related to or could have contributed to visual issues were observed.Based on the return condition of the lens (cut in half), no further product evaluation could be performed on the lens.No handpiece was received as part of this return therefore no product evaluation could be performed on the handpiece.Manufacturing record review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search revealed that no additional complaints were received for this production order.Conclusion: as a result of the investigation there is no indication of a product quality deficiency, and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13690773
Report Number3012236936-2022-00589
Device Sequence Number1869117
Product Code MFK
UDI-Device Identifier05050474751262
UDI-Public(01)05050474751262(17)240714
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/14/2024
Device Model NumberDFW150
Device Catalogue NumberDFW150U200
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/09/2022
Supplement Date Received by Manufacturer04/22/2022
Initial Report FDA Received Date03/08/2022
Supplement Report FDA Received Date04/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-