Model Number 1039370
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging an everflo oxygen concentrator had thermal damage to the power cord.There was no report of patient harm or injury.The device has yet to be returned to the manufacturer for evaluation.A follow up report will be submitted when the manufacturer has completed the investigation.
 
Additional Manufacturer Narrative
The manufacturer previously reported an everflo oxygen concentrator alleging thermal damage to the power cord.There was no report of patient harm or injury.The power cord was not returned to the manufacturer for evaluation.The manufacturer was provided photographs of the power cord.The manufacturer reviewed the photographs and determined the power cord had physical damage closest to the entrance into the enclosure of the device and on the power cord itself.The power cord had two openings that separated.The device has been constrained against a flat surface and the continuous bending has fatigued the wires to the point of failure.Product labeling states,"for proper operation, your concentrator requires unobstructed ventilation.The ventilation ports are located at the rear base of the device and at the side air inlet filter.Keep the device at least 15 to 30 cm away from walls, furniture, and especially curtains that could impede adequate airflow to the device.Do not place the concentrator in a small closed space (such as a closet)." do not use the oxygen concentrator if either the plug or power cord is damaged.Do not use extension cords or electrical adapters." the manufacturer concludes the damage to the power cord was caused by physical damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERFLO OXYGEN CONCENTRATOR
Common Device Name
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
PHILIPS RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
PHILIPS RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key13690964
Report Number2518422-2022-09184
Device Sequence Number17421139
Product Code CAW
UDI-Device Identifier00606959020436
UDI-Public00606959020436
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K061261
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1039370
Device Catalogue Number1039370
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/22/2022
Supplement Date Received by Manufacturer09/12/2022
Initial Report FDA Received Date03/08/2022
Supplement Report FDA Received Date09/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/28/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-