• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS SMARTSITE GRAVITY BLOOD SET; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS SMARTSITE GRAVITY BLOOD SET; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 47233E
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that 5 bd alaris smartsite gravity blood sets experienced kinked tubing.The following information was provided by the initial reporter: tubing is crimped and doesn't allow flow.
 
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.(b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 03-mar-2022.H6: investigation summary: a complaint of tubing being crimped and flow not passing through was received from the customer.One used and six unused samples were returned for investigation.The used sample had a kink in the tubing below the drip chamber.The customer complaint was confirmed.5 of the unused samples were kinked in the same location.Set was primed and air bubbles could be seen in the tubing.The air bubbles were removed and the set was then infused at 250 mll/hr.No issues were observed.A device history record review for model 47233e lot number 20125725 was performed.The search showed that a total of 9,603 units in 1 lot number was built on 09dec2020.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.The manufacturing plant was notified and an investigation was performed.The root cause for this issue was determined to be coiling and packaging set prior to the added solvent fully drying.H3 other text : see h10.
 
Event or Problem Description
It was reported that 5 bd alaris smartsite gravity blood sets experienced kinked tubing.The following information was provided by the initial reporter: tubing is crimped and doesn't allow flow.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS SMARTSITE GRAVITY BLOOD SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13691025
Report Number9616066-2022-00208
Device Sequence Number1945090
Product Code FPA
UDI-Device Identifier50885403235441
UDI-Public50885403235441
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K820278
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date12/09/2023
Device Model Number47233E
Device Catalogue Number47233E
Device Lot Number20125725
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2022
Supplement Date Received by Manufacturer04/04/2022
Initial Report FDA Received Date03/08/2022
Supplement Report FDA Received Date04/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/04/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-