Model Number GM505RA
Medical Device Problem Code Low Readings (2460)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/16/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
I-sens conducted a cause analysis of the reported problem with the returned meter.When testing the control solution with the returned meter and normal strips 10 times, the meter detected the strip insertion all 10 times but did not detect the application of the solution 7 times.Also, the lo message occurred in the remaining 3 times where the measurement was performed.As a result of disassembling the meter, it was confirmed that foreign substances had flowed into the connector.
 
Event or Problem Description
Customer reported issues of does not register solution/blood.Complaint was reviewed for investigation and reportability on 11/15/2021.Complaint was marked for investigation and determined not reportable per arkray's per internal procedures.On december 16, 2021 returned meter, failed testing by generating low results, out of specification.The meter gave low readings, making this complaint reportable to the manufacturer as a malfunction.December 16, 2021, marks the aware date for mdr filing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURE PRISM MULTI BLOOD GLUCOSE TEST SYSTEM
Common Device Name
BLOOD GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
I-SENS, INC.
43, banpo-daero, 28-gil
seocho-gu
seoul, 06646
KS  06646
Manufacturer (Section G)
I-SENS WONJU FACTORY
94-1, donghwagongdan-ro
munmak-eup
wonju-si, 26365
KS   26365
Manufacturer Contact
hee seon
43, banpo-daero, 28-gil
seocho-gu
seoul, 06646
MDR Report Key13691917
Report Number3005621222-2022-00001
Device Sequence Number1943684
Product Code NBW
UDI-Device Identifier00015482530013
UDI-Public(01)00015482530013
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K151164
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 03/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberGM505RA
Device Catalogue Number530001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/11/2022
Initial Report FDA Received Date03/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/18/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1