Model Number DAT015
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A nurse reported that during implantation of an intraocular lens (iol) there was scratch on the lens.There was patient contact.Additional information was requested.
 
Additional Manufacturer Narrative
The lens was returned in an opened non-company peel pouch.Solution was dried on the lens.The optic was torn/split/cracked and cut into multiple pieces, typical of insertion and removal.The returned cut optic portions have multiple scratches and deep scrape marks on the anterior and posterior sides of the lens.A large portion of cut optic lens material is missing (not returned).Qualified associated products were indicated.The viscoelastic indicated is not qualified for the product combination used.The root cause for the reported complaint appears to be related to failure to follow the ifu(instructions for use).The file indicated the use of a viscoelastic that is not qualified for the product combination used.The ifu instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information was requested and received stating there was no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ VIVITY EXTENDED VISION UV ABSORBING IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13692818
Report Number1119421-2022-00417
Device Sequence Number13929717
Product Code POE
UDI-Device Identifier00380652424691
UDI-Public00380652424691
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/30/2026
Device Model NumberDAT015
Device Lot Number15208158
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/16/2022
Supplement Date Received by Manufacturer03/09/2022
Initial Report FDA Received Date03/08/2022
Supplement Report FDA Received Date04/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HEALON; MONARCH III IOL INJECTOR; MONARCH III IOL CARTRIDGE D
Patient Age71 YR
Patient SexFemale
-
-