Model Number 2361-5065S
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/30/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported: "one month postoperative examination showed backout of the advanced locking screw.At the moment, no additional information is available and no revision is planned.".
 
Additional Manufacturer Narrative
Device is not accessible.If additional information becomes available, it will be provided in a supplemental report.
 
Additional Manufacturer Narrative
Please note the correction to h6 health impact code.The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.A device inspection was not possible since the affected device was not returned and no other evidences were provided for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.There was no information about the patient's post-op behavior.It could not be determined if the patient had contributed to the event.There were no x-rays available allowing a medical review.The item was still implanted, thus a physical investigation was not possible.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
It was reported: "one month postoperative examination showed backout of the advanced locking screw.At the moment, no additional information is available and no revision is planned.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED LOCKING SCREW IMN SCREWS Ø5X65MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13693463
Report Number0009610622-2022-00085
Device Sequence Number1944839
Product Code HWC
UDI-Device Identifier07613327381634
UDI-Public07613327381634
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K191271
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number2361-5065S
Device Catalogue Number23615065S
Device Lot NumberK07D05F
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/08/2022
Supplement Date Received by Manufacturer06/10/2022
Initial Report FDA Received Date03/08/2022
Supplement Report FDA Received Date06/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/07/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-