Product analysis conclusion: the reported type ia and ib endoleak could not be confirmed on films provided; therefore , the cause of the events could not be determined.Lack of pre-implant ct¿s did not allow for a thorough assessment of the pre-implant anatomy.The post implant ct's received were non-contrast; therefore, no assessment of any endoleak, stent graft patency or thrombus could be performed.Angiograms (including endoleak interrogation) could allow for a more comprehensive determinations of the endoleak source/cause.It was reported by the physician that anatomical considerations , such as dilation of the aortic neck and iliac arteries were the cause of the reported endoleaks.B5; as per the physician, the cause of the type ia and ib endoleaks is due to continued dilatation of the aortic neck and common iliac arteries.It was reported that the limb model model enlw1620c124e, sn (b)(6)implanted during the index procedure on the (b)(6) 2012 also contained an endoleak type ib, as per the physician.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|