• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. BASE PLATE UNCEMENTED; SHOULDER PROSTHESIS, REVERSE CONFIGURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. BASE PLATE UNCEMENTED; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Catalog Number 00434903811
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Localized Skin Lesion (4542); Skin Infection (4544)
Date of Event 02/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Device product code - phx.Concomitant products: item# 00434904011; lot# 64032723.Item# 00434904000; lot# 64560444.Item# 0104223036; lot# 3043248.Item# 0104223042; lot# 3021386.Report source: foreign - event occurred in (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product was sent for pathology testing.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that the patient was revised approximately one (1) year and one (1) month post-implantation due to screw fracture and loosening of baseplate.Additionally patient had an infected skin lesion treated three (3) weeks prior to revision.Surgeon mentioned that infected skin lesion and the fact the patient is a farmer with high physical demands may have contributed to the baseplate loosening.Attempts have been made and no further information has been provided.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Medical records were not provided.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.The sterile cert within the device history record was reviewed and found that the devices were sterilized per specification.A definitive root cause cannot be determined for loosening.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: the histopathology reported mild pale histiocytic infiltrate, suggestive of prosthesis wear; degenerative changes; no evidence of infection.The root cause of the previous investigation does not change.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BASE PLATE UNCEMENTED
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13693620
Report Number0001822565-2022-00687
Device Sequence Number4494530
Product Code PHX
UDI-Device Identifier00889024269101
UDI-Public(01)00889024269101(17)300430(10)64556002
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K130661
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number00434903811
Device Lot Number64556002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2022
Supplement Date Received by Manufacturer04/06/2022
05/13/2022
Initial Report FDA Received Date03/08/2022
Supplement Report FDA Received Date05/04/2022
05/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/18/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NI
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NARRATIVE IN H10
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age79 YR
Patient SexMale
Patient Weight90 KG
-
-