• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINNACLE MULTIHOLE II CUP 50MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINNACLE MULTIHOLE II CUP 50MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Model Number 1217-20-050
Medical Device Problem Codes Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Osteolysis (2377)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2012, the patient underwent the primary tha surgery with the cup, the stem and the head implants in question.After the surgery, it was confirmed that osteolysis suspected to be armd at the back of the cup on an unknown date.A revision surgery is planned on (b)(6) 2022.No further information is available.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Information was received, indicating that "it is noted, that the armd within the joint and around and behind the cup is suspected to be osteolysis.It is reasonable, to attribute the osteolysis to the metal on metal articulation of the metal head and metal liner.There is no noted, product deficiency related to the cup".Due to this, depuy synthes joint reconstruction considers, the device on this report to be not reportable, as there is no allegation of deficiency.Further updates will only be provided, if additional information is received, that changes the regulatory determination.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE MULTIHOLE II CUP 50MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13693649
Report Number1818910-2022-04240
Device Sequence Number20126401
Product Code LPH
UDI-Device Identifier10603295009610
UDI-Public10603295009610
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P070026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2022
Device Model Number1217-20-050
Device Catalogue Number121720050
Device Lot Number193934
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2022
Supplement Date Received by Manufacturer04/01/2022
05/18/2022
Initial Report FDA Received Date03/08/2022
Supplement Report FDA Received Date04/06/2022
05/20/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/01/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
9/10 ULTAMET 36MM HEADS +3; PINN CAN BONE SCREW 6.5MMX20MM; PINNACLE MTL INS NEUT36IDX50OD; SROM 9/10 18X12X135 36
Patient SexUnknown
-
-