(b)(4).The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.It was reported that the device was prepped inside the anatomy.It should be noted that the percutaneous transluminal angioplasty (pta), armada 14 xt, over the wire (ot), global instructions for use (ifu) specifies: all air must be removed from the balloon and displaced with contrast medium (diluted 1:1 with normal saline) prior to inserting into the body (repeat steps 4a through 4g, if necessary); otherwise, complications may occur.In this case, it is unknown if the ifu violation caused or contributed to the reported complaint.The investigation was unable to determine a conclusive cause for the reported balloon rupture.Balloon material ruptures can be affected by numerous factors including, but not limited to, balloon damage during processing of the balloon material, materials, inflation technique, and interactions with other devices, lesion calcification, or insufficient preparation prior to use.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|