WILLIAM A. COOK AUSTRALIA, PTY LTD ZENITH FENESTRATED AAA ENDOVASCULAR GRAFT PROXIMAL BODY; MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Model Number |
G32534 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Difficulty was experienced advancing the device into the aorta.The physician unsuccessfully attempted to dilate the iliacs.An attempt to access the iliac limbs from the other side was tried, but it was deemed to be less favourable.A conservative approach was then taken, and the device was removed.Finally, a self-expanding stent was placed in each iliac.Treatment options will be reviewed for further follow up/procedures at a later date.
|
| |
|
Additional Manufacturer Narrative
|
|
The undeployed device was returned for evaluation.It was observed to be contaminated with biological matter.The inner cannula was noted to be slightly bent.No damage was observed to the dilator tip or any other part of the device.No images were supplied to assist the investigation.Additional information was received that stated the following: the device was not implanted; however, they had trouble while trying to advance it.The procedure was not completed, but the patient did well.The work order for lot: ac1089720 was reviewed and appear complete and correct.The associated inspection records confirms that the device was manufactured to specification.The review of specifications found that there are number of processes and checks in place to ensure that any device deficiency would be identified prior to shipment.The instructions for use (ifu) supplied with the device was reviewed and states the following: access vessel diameter (measured inner wall to inner wall) and morphology (minimal tortuosity, occlusive disease, and/or calcification) should be compatible with vascular access techniques and delivery systems of the profile of a 14 french to 22 french vascular introducer sheaths.Iliac conduits may be used to ensure the safe insertion of the introduction system.Vessels that are significantly calcified, occlusive, tortuous or thrombus-lined may preclude placement of the endovascular graft and/or may increase the risk of embolization/trauma.Key anatomic elements that may affect successful exclusion of the aneurysm include severe proximal neck angulation (> 45 degrees for infrarenal neck to axis of aaa or > 45 degrees for suprarenal neck relative to the immediate infrarenal neck); short proximal aortic neck (<4 mm); greater than 10% increase in diameter over 15 mm of proximal aortic neck length; and circumferential thrombus and/or calcification at the arterial implantation sites, specifically the proximal aortic neck and distal iliac artery interface.Irregular calcification and/or plaque may compromise the fixation and sealing of the implantation sites.Necks exhibiting these key anatomic elements may be more conducive to graft migration.Based on the available information, inspection of the returned device and the completed investigation, the root cause was determined to be procedure related; due to patient conditions, the delivery system failed to access the aneurysm site.
|
| |
|
Event or Problem Description
|
|
Difficulty was experienced advancing the device into the aorta.The physician unsuccessfully attempted to dilate the iliacs.An attempt to access the iliac limbs from the other side was tried, but it was deemed to be less favourable.A conservative approach was then taken, and the device was removed.Finally, a self-expanding stent was placed in each iliac.Treatment options will be reviewed for further follow up/procedures at a later date.
|
| |
|