Model Number RBYLP0105
Medical Device Problem Codes Failure to Advance (2524); Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
The patient was undergoing a coil embolization procedure to treat a gastrointestinal bleed (gi) using ruby coil lps and a non-penumbra microcatheter.During the procedure, the ruby coil lp would not advance and it came apart.The procedure was completed using other ruby coil lps and the same microcatheter.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Additional Manufacturer Narrative
Evaluation of the returned ruby coil lp revealed that the pusher assembly mid-joint was retracted proximal to the introducer sheath friction lock.Evaluation revealed that the pusher assembly was fractured near the mid-joint.If the pusher assembly mid-joint is retracted proximal to the introducer sheath friction lock, resistance may be experienced during advancement.Forceful advancement against this resistance likely contributed to the kink and fracture observed near the pusher assembly mid-joint.Further evaluation revealed that the embolization coil was detached from its pusher assembly inside its introducer sheath.This likely resulted from the pusher assembly fracture, which resulted in the pull wire being retracted out of the distal fractured segment.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL LP
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key13694297
Report Number3005168196-2022-00108
Device Sequence Number13939113
Product Code HCG
UDI-Device Identifier00815948021730
UDI-Public00815948021730
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K192955
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/06/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRBYLP0105
Device Catalogue NumberRBYLP0105
Device Lot NumberF110453
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2022
Supplement Date Received by Manufacturer03/21/2022
Initial Report FDA Received Date03/08/2022
Supplement Report FDA Received Date04/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
Date Report Sent to FDA01/01/2005
-
-