• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES REAL INTELLIGENCE DRILL GUIDE; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES REAL INTELLIGENCE DRILL GUIDE; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10021
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during pin placement in a cori assisted tka surgery, the real intelligence drill guide was used and the pins bind up and got stuck.The procedure was finished with a smith and nephew back up device without delay.The patient was not harmed.
 
Additional Manufacturer Narrative
Internal complaint reference: case: (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the real intelligence drill guide, part number rob10021, used for treatment was not returned for evaluation.A relationship between the reported event and the device could not be established.A complaint history review for similar reported/confirmed complaints has identified prior events.While all products meet required manufacturing specifications prior to release a serial number, lot number, or part revision is required to link the device to a dhr or nc investigation.The failure mode and associated risk have been anticipated within the risk file.The risk level is still adequate.Although the reported problem was not confirmed through a visual or functional evaluation, factors contributing to the reported symptom may have been associated with improper technique to insert and secure bone pin.Based on the investigation, the need for a corrective action is not recommended or required at this time.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE DRILL GUIDE
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13694360
Report Number3010266064-2022-00175
Device Sequence Number19300065
Product Code OLO
UDI-Device Identifier00885556757406
UDI-Public00885556757406
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K193120
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberROB10021
Device Catalogue NumberROB10021
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/16/2022
Supplement Date Received by Manufacturer07/31/2022
Initial Report FDA Received Date03/08/2022
Supplement Report FDA Received Date08/04/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CORI ROBOTICS USA-(B)(4)
Patient SexUnknown
-
-