| Model Number |
PFSR100026 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: case: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during a navio assisted tka surgery, the navio handpiece thumberscrew was fractured.No piece fell inside the patient.The procedure was finished with a smith and nephew back up device without delay.The patient was not harmed.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the navio handpiece thumberscrew, part number pfsr100026, lot number 000000092, used for treatment was not returned for evaluation.A relationship between the reported event and the device could not be established.A complaint history review for similar reported/confirmed complaints has identified prior events.While all products meet required manufacturing specifications prior to release the provided product identification information (lot) did not match any known release of this part.The failure mode and associated risk have been anticipated within the risk file.The risk level is still adequate.Although the reported problem was not confirmed through a visual or functional evaluation, factors contributing to the reported symptom may have been associated with repeated over torquing resulting in thumbscrew shaft fracture.Based on the investigation, the need for a corrective action is not recommended or required at this time.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.Corrected data: h6 (health effect - impact code).
|
| |
|
Search Alerts/Recalls
|