BOSTON SCIENTIFIC CORPORATION GUIDEZILLA; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 1867
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/23/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by manufacturer: the device was returned for analysis.The device was visually and microscopically examined.There was blood present in the shaft of the device.The collar was partially detached.At the most distal end of the collar, the shaft was partially detached and torn.The ends of the detached shaft are damaged and slightly twisted.The jagged edges of the detached shaft indicated that the device was likely kinked and rotated with force during handling of the device.
 
Event or Problem Description
Reportable based on device investigation completed on 13feb2022.It was reported that the catheter was kinked.The 80% stenosed target lesion was located in the moderately tortuous and moderately calcified left anterior descending artery.A guidezilla guide extension catheter was selected for use.During the procedure, it was noted that the catheter was kinked.The procedure was completed with another of the same device.There were no complications reported and the patient is stable.However, device investigations revealed that the collar was partially detached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUIDEZILLA
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC SCIMED, INC
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13694437
Report Number2134265-2022-02164
Device Sequence Number5046009
Product Code DQY
UDI-Device Identifier08714729839859
UDI-Public08714729839859
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K123765
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/08/2022
Device Model Number1867
Device Catalogue Number1867
Device Lot Number0026150141
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/13/2022
Initial Report FDA Received Date03/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/08/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age70 YR
Patient SexFemale
Patient Weight67 KG
-
-