• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR OMNILINK ELITE PERIPHERAL STENT SYSTEM; PERIPHERAL STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR OMNILINK ELITE PERIPHERAL STENT SYSTEM; PERIPHERAL STENT DELIVERY SYSTEM Back to Search Results
Model Number 1012631-39
Medical Device Problem Codes Difficult to Remove (1528); Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Foreign Body In Patient (2687); Embolism/Embolus (4438)
Date of Event 02/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents.The reported patient effect of embolism is listed in the omnilink elite® vascular balloon-expandable stent system, electronic instructions for use (eifu) as a known patient effect of coronary stenting procedures.The investigation determined the reported difficulties and subsequent foreign body in patient appear to be related to the operational context of the procedure.A conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that the procedure was performed to treat a dissection in the distal right iliac artery.An attempt was made to advance a sheath to the target site via left access; however, the sheath could not be advanced to the right distal iliac.The 8.0 x 39 mm omnilink elite stent delivery system was advanced; however, as there was no support from the sheath, the omnilink elite was unable to get to the target site.The device was removed; however, when pulling the stent into the sheath, the stent caught on the sheath and dislodged, remaining in the sheath.The stent delivery system was removed without difficulty.The stent remained in the sheath and an attempt was made to remove the sheath and stent.During removal, the stent embolized to the left profunda.A second physician was consulted, and the decision was made to leave the stent where it was, as the patient did not experience any adverse patient effect.The stent remained unexpanded in the left profunda and no additional intervention was performed.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNILINK ELITE PERIPHERAL STENT SYSTEM
Common Device Name
PERIPHERAL STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13694505
Report Number2024168-2022-02338
Device Sequence Number14730151
Product Code NIO
UDI-Device Identifier08717648178511
UDI-Public08717648178511
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
P110043
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2022
Device Model Number1012631-39
Device Catalogue Number1012631-39
Device Lot Number9052041
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/14/2022
Initial Report FDA Received Date03/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/20/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age75 YR
Patient SexMale
Patient Weight70 KG
-
-