| Model Number |
2088TC/100 |
| Medical Device Problem Codes |
Capturing Problem (2891); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/22/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported the patient presented in clinic for follow up.Chest x-rays taken showed the atrial lead had dislodged.The atrial lead was explanted and replaced.There were no patient consequences.
|
| |
|
Event or Problem Description
|
|
New information noted the atrial lead exhibited high capture thresholds.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported events of lead dislodgement and high capture thresholds were not confirmed.A partial lead was returned in one piece for analysis.Helix and electrical analysis were normal with no anomalies found.
|
| |
|