• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012276-15
Medical Device Problem Codes Deflation Problem (1149); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was to treat a mildly tortuous, 60% stenosed lesion in the left anterior descending (lad) artery.An unspecified stent was implanted and a 3.5x15mm trek rx balloon dilatation catheter (bdc) was advanced and used for post-dilatation without issue.The balloon was inflated 2 times to 10 atmospheres (atm).During deflation of the trek balloon, the physician believed that it took too long to deflate the balloon, however it did fully deflate, and it is unknown how long negative was held for during deflation.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was received.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
 
Event or Problem Description
Subsequent to the initial filed report, it was noted that the contrast mix was 30 percent contrast and 70 percent saline.No additional information was provided.
 
Additional Manufacturer Narrative
H6: medical device problem code 2017 - incorrect contrast mix.Visual, dimensional and functional inspections were performed on the returned device.The reported deflation issue was not confirmed.There were no anomalies noted during deflation and the balloon was folded in trifold state after each deflation, thus the reported complaint was not confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.It was reported that the contrast mix used during the procedure was 30 percent contrast and 70 percent saline.It should be noted that the trek rx coronary dilatation catheters (cdc), global instructions for use (ifu) specifies that the contrast is 60% contrast medium diluted 1:1 with normal saline.In this case, since the contrast ratio used was less than the required percentage it is unknown if the ifu violation caused or contributed to the reported complaint.The investigation was unable to determine a conclusive cause for the reported deflation issue.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13694616
Report Number2024168-2022-02344
Device Sequence Number13920480
Product Code LOX
UDI-Device Identifier08717648138577
UDI-Public08717648138577
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K103110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2023
Device Model Number1012276-15
Device Catalogue Number1012276-15
Device Lot Number00604G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/14/2022
Supplement Date Received by Manufacturer03/18/2022
Initial Report FDA Received Date03/08/2022
Supplement Report FDA Received Date04/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/04/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-