DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Output below Specifications (3004)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/10/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The full name of the event site was shortened due to field character limit; the full name is (b)(6) medical center.Communications/ interviews: biomed was assisted in printing out an alarm history strip to review all alarms.After a quick review, the biomed realized the pump did not alarm ¿constantly¿ with different issues.The helium gauge showed the tank was half full while the icon says ¿low¿ in red.Biomed was assisted to open the tank fully and to go by the gauge versus the computer screen icon.Cs explained how many auto-fills they would have once the gauge showed ¿in the red¿.Two helium tanks were provided by the customer to have bedside while the patient continued therapy.A supplemental report will be submitted upon completion of our investigation.Not returned to manufacturer.
|
|
Event Description
|
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) gave a low helium alarm.No patient harm, serious injury or adverse event was reported.
|
|
Manufacturer Narrative
|
At this time, the customer has not requested getinge to evaluate the iabp.Additional information was requested from the customer with regard to the repair and status of the iabp; however, despite our best efforts, no repair information and no status of the iabp has been received.If any pertinent information is received in the future, a supplemental report will be submitted.
|
|
Event Description
|
N/a.
|
|
Search Alerts/Recalls
|
|
|