• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Output below Specifications (3004)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/10/2022
Event Type  malfunction  
Manufacturer Narrative
The full name of the event site was shortened due to field character limit; the full name is (b)(6) medical center.Communications/ interviews: biomed was assisted in printing out an alarm history strip to review all alarms.After a quick review, the biomed realized the pump did not alarm ¿constantly¿ with different issues.The helium gauge showed the tank was half full while the icon says ¿low¿ in red.Biomed was assisted to open the tank fully and to go by the gauge versus the computer screen icon.Cs explained how many auto-fills they would have once the gauge showed ¿in the red¿.Two helium tanks were provided by the customer to have bedside while the patient continued therapy.A supplemental report will be submitted upon completion of our investigation.Not returned to manufacturer.
 
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) gave a low helium alarm.No patient harm, serious injury or adverse event was reported.
 
Manufacturer Narrative
At this time, the customer has not requested getinge to evaluate the iabp.Additional information was requested from the customer with regard to the repair and status of the iabp; however, despite our best efforts, no repair information and no status of the iabp has been received.If any pertinent information is received in the future, a supplemental report will be submitted.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key13694969
MDR Text Key288860889
Report Number2249723-2022-00480
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 02/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/10/2022
Initial Date FDA Received03/08/2022
Supplement Dates Manufacturer Received03/08/2022
02/01/2024
Supplement Dates FDA Received04/04/2022
02/02/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/01/2013
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
MAQUET BALLOON.
-
-