Model Number MS1-4570S
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
 
Event or Problem Description
Information was received that the rod failed to distract.No patient adverse event was reported.
 
Event or Problem Description
No additional information was provided.
 
Additional Manufacturer Narrative
Device evaluation: upon receipt of the returned product score marks consistent with incremental distraction on the distraction rod.The distraction rod was able to be lengthened by hand which indicated a broken locking pin.X-ray images of the internal components and sectioning the rod confirmed the locking pin was broken.Sectioning the rod additionally revealed corrosion around the looking pin.As a result of the locking pin breakage, there is nothing constraining the axial motion of the distraction rod within the housing tube; this is usually known as "pistoning" which confirmed the reported failure.The root cause is likely due to corrosion and bending stresses.A capa was completed in which a new pin was implemented with 65 percent improved strength and significantly improved corrosion resistance.Device history records (dhr) review revealed that this unit was built prior to the release of the new pin.Device records review: a review of the device history records (dhr) confirmed the unit met all of the required inspections and specifications prior to release.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Common Device Name
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS
101 enterprise #100
aliso viejo 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS
101 enterprise #100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise #100
aliso viejo, CA 92656
8583448112
MDR Report Key13695943
Report Number3006179046-2022-00140
Device Sequence Number3905491
Product Code PGN
UDI-Device Identifier00812258026332
UDI-Public812258026332
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
K201543
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMS1-4570S
Device Lot NumberA150304-10-00-02U
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/16/2022
Supplement Date Received by Manufacturer04/07/2022
Initial Report FDA Received Date03/08/2022
Supplement Report FDA Received Date04/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/04/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
Patient Weight27 KG
-
-