Model Number 370892-03
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) received the endoscope involved with this complaint and completed the device evaluation.Failure analysis investigation confirmed the reported complaint.The endoscope had mechanical damage to the distal window hermetic seal resulting in fluid invasion and subsequent major damage to optical components.The endoscope also had mechanical damage to laser marking and/or deep scratches to the ocular sleeve and shaft assembly.Fragments of adhesive of the distal window were missing.A review of the site's complaint history does not show any additional complaints related to this product and/or this event.Verification of the si endoscope product via system logs cannot be performed because si endoscope product details are not captured in the system log.No image or video was provided for review.This complaint is being reported due to the following conclusion: failure analysis acknowledges that a fragment was missing from a portion of the device that enters the patient.Based on the information provided, there was no report from the customer that a fragment from the instrument fell into the patient during the procedure.While there was no report of harm or injury to the patient and no report of fragments falling into the patient, the reported failure mode could likely cause or contribute to an adverse event if the fragment(s) did fall into the patient.
 
Event or Problem Description
It was reported that during a da vinci-assisted surgical procedure, the endoscope had a fogging issue.There was no reported injury.Intuitive surgical, inc.(isi) followed up with the initial reporter and obtained the following additional information: there was no record of patient injury.The procedure was completed with a backup endoscope.The operation was prolonged about 5 to 10 minutes.The reporter did not have the requested patient information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DA VINCI
Common Device Name
ENDOSCOPE, FIREFLY
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key13696220
Report Number2955842-2022-10533
Device Sequence Number20272785
Product Code NAY
UDI-Device Identifier00886874110201
UDI-Public(01)00886874110201
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K101077
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 02/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number370892-03
Device Catalogue Number370892
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/09/2022
Initial Report FDA Received Date03/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES
Patient SexUnknown
-
-