| Model Number |
07P5121 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer obtained false positive alinity i total b-hcg results for two samples.The following initial and repeat results were provided: (b)(6) 2022 patient 1: 56.7 and <2.4 miu/ml ,(b)(6) 2022 patient 2: 196.0 and <2.4 miu/ml , and no impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.Section a patient information: no specific patient information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
After further evaluation, the suspect medical device was changed from alinity i b-hcg, list number 07p51-21 and manufacturing site longford, irl to alinity i processing module, list number 03r65-01, and manufacturing site irving, tx.Mdr number 3016438761-2022-00218 has been submitted and all further information will be documented under that mdr number.H3 other text : the cause of the issue was determined to be the analyzer refer to mdr # 3016438761-2022-00218.
|
| |
|