Model Number TRIEEAXL28MT
Medical Device Problem Code Misfire (2532)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during a robotic laparoscopic low anterior resection, on anastomosis, there was a torn or incomplete proximal tissue donut.There was no patient injury.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a robotic laparoscopic low anterior resection, on anastomosis, there was a torn or incomplete proximal tissue donut.It was noted that the anastomosis was intact and no air bubble observed when surgeon performs routinely air test leak and flexible sigmoidoscopy.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the donut formed poorly or was missing completely after firing.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EEA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
tracy landers
5920 longbow drive
boulder, CO 80301
3035816943
MDR Report Key13697665
Report Number2647580-2022-00758
Device Sequence Number6440219
Product Code GDW
UDI-Device Identifier10884521593121
UDI-Public10884521593121
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K172361
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2025
Device Model NumberTRIEEAXL28MT
Device Catalogue NumberTRIEEAXL28MT
Device Lot NumberP0H0458Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2022
Supplement Date Received by Manufacturer03/08/2022
05/17/2022
Initial Report FDA Received Date03/08/2022
Supplement Report FDA Received Date03/24/2022
05/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/26/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1