• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. SJM MASTERS SERIES HEMODYNAMIC PLUS VALVE; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. SJM MASTERS SERIES HEMODYNAMIC PLUS VALVE; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 15AHPJ-505
Medical Device Problem Code Incomplete Coaptation (2507)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2021 a patient with congenital aortic stenosis and epstein anomaly underwent a implant procedure of a 15mm sjm masters series hemodynamic plus valve.On (b)(6) 2021 structural valve deterioration was reported.One of the posterior leaflets was not moving.The patient was discharged on (b)(6) 2021.The leaflet was still not moving at the time of discharge.No additional information has been provided.11feb22 pm.(b)(4) - 15mm surgical aortic valve pas, (b)(4).
 
Additional Manufacturer Narrative
An event of structural valve deterioration of a 15mm masters series hemodynamic plus valve was reported.A more comprehensive assessment could not be performed as the device remains implanted and was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM MASTERS SERIES HEMODYNAMIC PLUS VALVE
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
177 east county road b
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
177 east county road b
st. paul MN 55117
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key13697906
Report Number3005334138-2022-00213
Device Sequence Number12353139
Product Code LWQ
UDI-Device Identifier05415067016515
UDI-Public05415067016515
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Study,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/08/2024
Device Model Number15AHPJ-505
Device Catalogue Number15AHPJ-505
Device Lot Number7037598
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2022
Supplement Date Received by Manufacturer04/12/2022
Initial Report FDA Received Date03/08/2022
Supplement Report FDA Received Date04/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/10/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age17 MO
Patient SexMale
Patient Weight5 KG
-
-