Model Number B2120-060
Medical Device Problem Codes Material Rupture (1546); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
Patient id: (b)(6).It was reported that on (b)(6) 2022, the patient presented with a left iliac vein, 100% restenosed lesion that had been previously treated with an unspecified stent.Balloon angioplasty was successfully performed using two armada 35 catheters.Two additional armada catheters were successfully advanced.Both device balloons had ruptured below rated burst pressure.The ruptures were suspected due to the previously implanted stent struts of the unspecified stent.The devices were successfully withdrawn.There was no adverse patient effect and no clinically significant delay.No additional information was provided regarding this issue.
 
Additional Manufacturer Narrative
Medical device problem - (b)(4).The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.In this case, it was reported that the balloon was inflated one time to 10 atmospheres.It should be noted that the armada 35 instruction for use (ifu) states: ¿inflation in excess of the rated burst pressure may cause the balloon to rupture.Use of a pressure monitoring device is recommended.¿ the rated burst pressure for this device is 9 atmospheres as indicated on the product label.It could not be determined if inflating the balloon slightly over the rated burst pressure contributed to the rupture.The investigation determined the reported balloon rupture appears to be related to operational context.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.The other armada catheter referenced is filed under a separate medwatch report number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13698725
Report Number2024168-2022-02367
Device Sequence Number5434532
Product Code LIT
UDI-Device Identifier08717648155567
UDI-Public08717648155567
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Model NumberB2120-060
Device Catalogue NumberB2120-060
Device Lot Number10906G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/28/2022
Initial Report FDA Received Date03/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age53 YR
Patient SexFemale
Patient Weight105 KG
-
-