• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION NC QUANTUM APEX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION NC QUANTUM APEX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 6533
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/09/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in the calcified descending artery.During the procedure, two 20mm x 2.50mm and a 20mm x 3.00mm nc quantum apex balloon catheters were used to dilate the lesion.However, during inflation at less than 14 atmospheres, the balloons ruptured.The devices were removed and upon checking the artery, the lesion was not dilated.A rotational atherectomy with a 1.75mm balloon was used to prepare the artery for stent implantation.No patient complications were reported and the patient status was stable.
 
Additional Manufacturer Narrative
The returned product consisted of an nc quantum apex balloon catheter.The device was visually and microscopically examined.There was contrast and blood in the inflation lumen and the balloon.The balloon was loosely folded.The device was prepped with an inflation device filled with water and connected to the inflation port to inflate the device.At 13mm from the tip of the device there was a pinhole in the balloon.The device failed to inflate to rated burst pressure.
 
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in the calcified descending artery.During the procedure, two 20mm x 2.50mm and a 20mm x 3.00mm nc quantum apex balloon catheters were used to dilate the lesion.However, during inflation at less than 14 atmospheres, the balloons ruptured.The devices were removed and upon checking the artery, the lesion was not dilated.A rotational atherectomy with a 1.75mm balloon was used to prepare the artery for stent implantation.No patient complications were reported and the patient status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC QUANTUM APEX
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key13699670
Report Number2134265-2022-02546
Device Sequence Number20322689
Product Code LOX
UDI-Device Identifier08714729783480
UDI-Public08714729783480
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K121667
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/08/2024
Device Model Number6533
Device Catalogue Number6533
Device Lot Number0027800955
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/25/2022
Supplement Date Received by Manufacturer08/01/2022
Initial Report FDA Received Date03/08/2022
Supplement Report FDA Received Date08/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-