ETHICON ENDO-SURGERY, LLC. ENDOPATH*THORACIC TROCAR SLEEV; CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number TT012
Medical Device Problem Codes Crack (1135); Component or Accessory Incompatibility (2897)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a semi-open pneumonectomy, a stapler was inserted/removed through the trocar.It got caught in the sleeve and sleeve was broken and fell inside the patient.The broken pieces were retrieved by forceps.The procedure was not converted.The device was used on the lung.Another device (trocar) was used to complete the case.There were no adverse consequences to the patient.The patient¿s condition is stable.No further information is available.
 
Additional Manufacturer Narrative
(b)(4).Batch # unk.Additional information received: a photo was received for review.Review is in progress and has not yet been completed.Upon completion of photo review, a supplemental medwatch will be sent.The lot/batch was not provided; therefore, a manufacturing record evaluation could not be performed.Attempts are being made to retrieve the device.To date, no device has been returned.If the device or further details are received at a later date, a supplemental medwatch will be sent.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
(b)(4).Date sent; 4/5/2022.A photo along with the device was submitted to ethicon for evaluation.During the visual analysis of the photo, the following was observed: the photo shows a sleeve cracked from the tip side.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that only the sleeve from the tt012 device was received with the tip broken and a pieces inside a plastic bag.The event reported was confirmed and it is related to improper use of the device.One possible cause for the damage found on the sleeve may be excessive external load placed on the device.Please reference the instruction for use for more information.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Additional information was requested, and the following was obtained: was there only difficulty inserting and removing echelon stapler? or was there some other issue with the echelon stapler? the echelon got caught in the trocar sleeve and it was broken and fell inside the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 
6107428552
MDR Report Key13701204
Report Number3005075853-2022-01413
Device Sequence Number5434016
Product Code GEA
UDI-Device Identifier10705036013372
UDI-Public10705036013372
Combination Product (Y/N)N
PMA/510(K) Number
K920110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTT012
Device Catalogue NumberTT012
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/08/2022
Supplement Date Received by Manufacturer03/09/2022
Initial Report FDA Received Date03/08/2022
Supplement Report FDA Received Date04/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-