BOSTON SCIENTIFIC CORPORATION EXPRESS SD RENAL/BILIARY; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 20188
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.A 6.0mmx18mmx150cm express sd renal/biliary stent was advanced for treatment.However, during the procedure, the shaft of the stent got buckled and snapped.The procedure was completed with a different device.There were no patient complications nor injuries reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr: returned product consisted of an express sd stent catheter.The outer shaft, inner shaft, balloon, and tip were visually and microscopically examined.Visual examination revealed kinks 12.3cm, 12.4 cm and 12.9 from the distal end.Microscopic examination revealed no other damage than the kinks.Product analysis did not confirm the complaint of a break.
 
Event or Problem Description
It was reported that shaft break occurred.A 6.0mmx18mmx150cm express sd renal/biliary stent was advanced for treatment.However, during the procedure, the shaft of the stent got buckled and snapped.The procedure was completed with a different device.There were no patient complications nor injuries reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESS SD RENAL/BILIARY
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13702723
Report Number2134265-2022-02087
Device Sequence Number5332422
Product Code NIN
UDI-Device Identifier08714729484844
UDI-Public08714729484844
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K152607
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/11/2024
Device Model Number20188
Device Catalogue Number20188
Device Lot Number0027295196
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/15/2022
Supplement Date Received by Manufacturer03/18/2022
Initial Report FDA Received Date03/09/2022
Supplement Report FDA Received Date04/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-