| Model Number |
364499 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.
|
| |
|
Event or Problem Description
|
|
An orsiro drug-eluting stent system was selected for treatment of the om lesion.During advancing the device, the stent dislodged inside the left main artery.The stent was retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes the returned product was subjected to a detailed technical analysis and the corresponding product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation revealed that the balloon is well folded and shows no signs of inflation.Microscopic inspection showed stent imprints on the exposed balloon surface, indicating that the stent was properly crimped in between the two radiopaque markers at the time of delivery.The stent was returned separately wrapped in an op cloth.The stent is severely deformed at both ends, i.E.Several struts are bent.Review of the product release documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.Based on the conducted investigations, no manufacturing or material related root cause was determined.
|
| |
|
Search Alerts/Recalls
|